Recall StatusTerminated
Date Initiated2020-07-15
Date Terminated2021-03-22
Product Quantity1474 kits
Reason for Recall
There is an increased rate of potential false positive results for the Influenza A (no subtype) target, which may lead to 1) inappropriate antiviral use, 2) erroneous or delayed differential diagnosis and missing of an alternative diagnosis.
Firm Action
On July 15, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers to inform them of the increased rate of potential false positive results for the Influenza A (no subtype) target.
Customers were instructed to do the following: l
-Please review and reassess all Influenza A (no subtype) positive samples tested with the affected lots.
-Please note that as per Instructions for Use if only an Influenza A signal is present and no additional
signal for any of the subtypes is generated, it can be due to either low concentration or, in very rare
cases, a new variant or any Influenza A strain other than H1 and H3 (e.g., H5N1, which can infect
humans).
Such result should, therefore, be interpreted in the context of epidemiological and clinical plausibility.
-If you have used the affected lot/s and you have concerns about previously generated Influenza A (no
subtype) positive results on the basis of which patient treatment/management was initiated, we
recommend a review of the corresponding Influenza A (no subtype) positive results.
-For continued use of the affected QIAstat-Dx Respiratory SARS-CoV-2 Panel REF 691223 LOTs, Flu A (no
subtype) positive results should be retested with an alternate method and interpreted in the context of
current epidemiology and clinical presentation.
-Review this notice with your laboratory/medical director.
-IMPORTANT: Forward this information to all individuals and departments within your organization using
the above listed kits. If you are not the end user, please forward this notice to the product end user.
-Please complete Acknowledgement of Receipt attached to this letter by 29 July 2020.
If you have any questions or concerns, please contact your local QIAGEN Technical Service
Department through any of the following:
Telephone: 800 362 7737
Email:
[email protected]
Root Cause
Material/Component Contamination
Distribution Pattern
US Nationwide distribution including in the states of AK, AL, CA, DC, DE, FL, GA, IL, IN, KY, MA, MD, MN, NC, NJ, NY, OH, OR, PA, TX, WA.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.