Medcomp 15.SF X 24CM Titan HD CATHETER (CUFF 19CM FROM TIP) Ref: THD155024
FDA Recall Z-2940-2020 · Posted None
Advertisement
Recalling FirmMedical Components, Inc dba MedComp
Recall StatusTerminated
Product CodeMSD
Date Initiated2020-07-27
Date Terminated2021-04-30
Product Quantity132 eaches
Reason for Recall
Kits were packaged with the incorrect size valve peelable introducer. The kits include a 15F instead of the correct 16F valve peelable introducer
Firm Action
Medcomp issued Urgent Medical Device Recall letter issued first class mail to the customers on July 27, 2020. Letter states reason for recall, health risk and action to take: " Immediately examine your inventory and quarantine all product subject to recall.
" Options for disposition of affected product:
o Return the product to Medcomp - Contact your Medical Components, Inc.
customer service representative for a Returned Goods Authorization {RGA)
number. The phone number to call is 215-256-4201.
o Have individually packaged and sterilized 16F peelable introducers (MR191603) sent to use with the kits. Call your customer service representative and provide the number of kits you need the replacements introducers for. Same number as above.
" Complete and return the Medical Device Recall Product Disposition Form. The form can be
returned by fax {215-256-9191) or email ([email protected]).
Contact Information for questions: Monday through Friday, 8:00 AM to 4:00 PM, Eastern Standard Time. Phone: 215-256-4201; Email: [email protected].
Root Cause
Under Investigation by firm
Distribution Pattern
AZ, CA. FL. GA, NC, ND, NJ, NY, OH, PA, VA, WA
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.