Recall StatusOpen, Classified
Date Initiated2024-07-01
Product Quantity100 units
Reason for Recall
Due to packaging defects, sterility assurance and heparin activity may be compromised.
Firm Action
Between the dates of July 18-19,2024 , W.L. Gore issued a "Urgent Medical Device Correction" notification to affected consignees either via UPS or hand delivery. W.L. Gore asked consignees to take the following actions:
1. For US Only, per the IFU, carefully inspect the foil pouch to ensure that it is not
compromised. Do not use the GORE VIABAHN VBX Balloon Expandable Endoprosthesis if the foil pouch is compromised. Do not use the GORE VIABAHN VBX Balloon Expandable Endoprosthesis if the sterile pouch is compromised. This warning includes the Compliance Chart embedded in the seal or a damaged/cut Compliance Chart.
2. Open the foil pouch, and gently remove the coil containing the Reduced Profile GORE
VIABAHN VBX Balloon Expandable Endoprosthesis and the Compliance Chart. Carefully
inspect the Compliance Chart to ensure that it is not damaged. Do not use the Reduced
Profile GORE VIABAHN VBX Balloon Expandable Endoprosthesis if the Compliance Chart cannot be removed from the pouch or is damaged as the Compliance Chart may have been embedded in the seal. Reference sample images below.
3. Implanting Physicians: If you suspect you have implanted a potentially compromised
Reduced Profile GORE VIABAHN VBX Balloon Expandable Endoprosthesis device, please
be mindful of the potential health outcomes and plan to take the needed measures that
could minimize any potential health hazard/risk/harms associated with a compromised
sterile barrier.
4. Please complete and sign the enclosed ACKNOWLEDGEMENT FORM and return to
[email protected] within 2 weeks of receipt of this notification. This notice needs to be shared with those who should be aware within your institution or to any organization where potentially affected devices have been transferred (as appropriate).
On 10/25/24, additional correction notices were sent to customers informing them that additional devices were found to be affected.
Root Cause
Package design/selection
Distribution Pattern
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of Japan.
Official Source