SICAT IMPLANT V2.0
FDA Recall Z-2951-2020 · Posted None
Advertisement
Recalling FirmSICAT GMBH & CO. KG
Recall StatusTerminated
Product CodeLLZ
Date Initiated2020-07-15
Date Terminated2024-01-03
Product Quantity853 units
Reason for Recall
A dentist found implant positions are not correctly exported from the implant planning software SICAT IMPLANT V2.0 - for the specific export format CMG.DXD.
Firm Action
On 07/15/2020 customer notification letter was sent to all users/customers who have the 2.0 version of the software installed in the system. These users are instructed to only use the system for specific cases as defined in the field safety notice. The firm plans on implementing a new software version in July 2020 to correct the issues found.
Root Cause
Software design
Distribution Pattern
Worldwide distribution