Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an injector head. , SHORT CEIL MNT, STD HORZ ARM, Catalog # OCS 115AS, Material Number 59880288
FDA Recall Z-2961-2018 · Posted None
Advertisement
Recalling FirmBayer Medical Care, Inc.
Recall StatusTerminated
Product CodeJAK
Date Initiated2018-06-28
Date Terminated2018-11-26
Product Quantity1
Reason for Recall
The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
Firm Action
On June 28, 2018, Urgent Medical Device Field Safety Corrective Action letters were distributed to customers. The letters informed the customers of the following:
A Bayer Service Representative will be scheduling a service visit to replace the ceiling column assembly of your OCS system(s) at no cost to you.
Root Cause
Equipment maintenance
Distribution Pattern
The devices were distributed to the following US states: OH, TX, and WV.