Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Foundation Modular Femoral Stem, Part Number, 220-10-110. Knee prosthesis component. The Foundation Revision Knee is a total knee system indicated as knee joint replacement for patients suffering from pain and dysfunction.

FDA Recall Z-2967-2018 · Posted None

Recalling FirmEncore Medical, Lp
Recall StatusTerminated
Product CodeJWH
Date Initiated2018-05-14
Date Terminated2020-08-17
Product Quantity12

Reason for Recall

After receipt of a product complaint for the labeling, it was determined that this device does not have FDA clearance.

Firm Action

The firm, DJO, sent an "URGENT FIELD SAFETY NOTICE" letter dated 5/14/2018 to it's customers via email, on May 14, 2018 and contained the following information: Description of Issue, Quantity in Field, product information and code, Action required, and an Acknowledgement form with contact information. The customers were instructed to: check your inventory for part number 220-10-110; Customer Service has already requested that these be quarantined by your agency; Contact Customer Service to request an RMA for return of these devices. As there are no currently cleared 10x100 femoral stem extensions, there is no product available for replacement; and complete the attached acknowledgment form, even if you do not have the affected product, and email the form to [email protected] by May 18, 2018. Modular Femoral Cemented Stem -- 220-10-110 -determined that this device does not have FDA clearance. If you have any questions, please contact Manager, Regulatory Affairs at (512) 834-6255 or by email at [email protected].

Root Cause

No Marketing Application

Distribution Pattern

U.S. Distribution to states of: AZ, MO, MS, NY, OK, and UT; internationally to: Japan.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑