ARTISTE/RT Therapist v4.2 and/or Oncologist v4.2 systems. Manufactured by Siemens AG, Medical Solutions, Kemnath, Germany. Delivery of X-ray photon and electron radiation for the therapeutic treatment of cancer.

FDA Recall Z-2977-2011 · Posted 2011-08-10

Recall StatusTerminated
Product CodeIYE
Date Initiated2011-06-03
Date Terminated2012-03-30
Product Quantity121 devices

Reason for Recall

A potential safety rick related to transferring reference images for offset calculation between a syngo RT Therapist v 4.2 or a syngo RT Oncologist 4.2 and other versions of the COHERENCE/syn go RT Therapist or Oncologist or vice versa. An incorrect offset may be applied for patient positioning and result in dose being delivered to the wrong location.

Firm Action

Siemens sent an "URGENT: MEDICAL DEVICE CORRECTION/CUSTOMER SAFETY ADVISORY NOTICE" letter dated June 3, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter provides preventive measure instructions for customers to use until an update for the system is available.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution-USA (nationwide) including the states of CA, FL, NC, NY, OH, PA, SD, UT, and WI and countries of Australia, Belgium, Croatia, Czech Republic, Germany, India, Ireland, Japan, Poland, Saudi Arabia, South Africa, Spain, Sweden, and United Kingdom.