Elecsys Anti-TPO Assay on cobas e analyzers 411, 601, 602, Catalog Number 06368590190
FDA Recall Z-2981-2020 · Posted None
Advertisement
Recalling FirmRoche Diagnostics Operations, Inc.
Recall StatusTerminated
Product CodeJZO
Date Initiated2020-08-20
Date Terminated2021-10-22
Product Quantity11,870 reagent kits
Reason for Recall
Recent internal verification studies found the current RF interference claim to no longer be met for the assay on the cobas e 411, 601, and 602 analyzers. Therefore, the claim has been updated.
Firm Action
Urgent Medical Device Correction notification letters dated 8/20/20 were distributed to customers.
Actions Required
" Be aware of the new RF interference claims listed in this Urgent Medical Device Correction (UMDC). This UMDC serves as labeling until the updated method sheet (V3), anticipated in September 2020, is available.
" Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. Customers do not need to perform an additional RF test unless otherwise determined in consultation with the physician or pathologist at your facility.
" Complete the attached fax back form (TP-01048) and fax or email it according to the instructions on the form.
" File this UMDC for future reference.
Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients.
The Elecsys Anti-TPO method sheet is being updated with the new claim for Rheumatoid Factor (RF) interference. Until the updated method sheet (V3) is available, this Urgent Medical Device Correction serves as your labeling. When these method sheets are updated, they will be available on the diagnostics.roche.com website and on the cobas e-library.
Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide