AIRO Mobile CT System Model # MobiCT-32
FDA Recall Z-2991-2018 · Posted None
Advertisement
Recalling FirmMobius Imaging, LLC
Recall StatusTerminated
Product CodeJAK
Date Initiated2018-08-01
Date Terminated2019-12-13
Product Quantity147
Reason for Recall
The Tube Current Modulation feature (Modulated Scans) is not working in AIRO systems with software version 2.0.0.0, and operators would not be able to detect this fault until after a scan is completed.
Firm Action
Mobius Imaging, LLC sent an URGENT MEDICAL DEVICE RECALL letter to customers advising that a voluntary recall related to the AIRO Mobile CT System was initiated. The letter included the reason for the recall and actions to be taken by the customers.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US and foreign distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement