Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT); 3) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT); 4) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT); 5) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT); 6) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL HTD, DUAL LIMB,COLUMN, Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT); 7) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD SINGLE LIMB,INFANT, Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT); 8) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD,INFANT,HTD,COLUMN,DRN, Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT); 9) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT); 10) MEDLINE HUDSON RCI,ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT); 11) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT).
FDA Recall Z-2993-2026 · Posted 2026-08-21
Advertisement
Recalling FirmMedline Industries, LP
Recall StatusOpen, Classified
Product CodeOGG
Date Initiated2026-07-13
Product Quantity271406 units
Reason for Recall
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Firm Action
Medline issued an Urgent Medical Device notice it its consignees on 7/13/2026 via email and USPS. The notice explained the issue, potential risk, and provided actions and precautions for healthcare providers.
If evidence of connector overheating or thermal damage is observed:
" Assess the patient and ensure continued respiratory support.
" Replace the affected circuit as soon as clinically appropriate.
" Remove the affected circuit from service and retain it for investigation."
For questions, contact the Recall Department at 866-359-1704 or [email protected].
Follow up notice sent on 7/27/2026 to clarify the item numbers.
***
Updated customer communication was sent out on 8/5/26. Consignees were instructed to destroy affected inventory. The firm has begun distributing corrected product.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide distribution - US Nationwide and the country of Panama.