ARTISTE Linear Accelerator with RT Therapist v4.1 system. Siemens Medical Solutions, Kemnath, Germany. Distributed by Siemens Medical Solutions, Concord, CA 94520. Product Usage: Linear Accelerators used to deliver X-ray photon and electron radiation for therapeutic treatment of cancer.

FDA Recall Z-3003-2011 · Posted 2011-08-11

Recall StatusTerminated
Product CodeIYE
Date Initiated2011-06-03
Date Terminated2012-02-17
Product Quantity223 active devices

Reason for Recall

A potential safety issue was discovered when a pause for an imaging segment is inserted (where an imaging segment does not exist) with the ARTISTE and RT Therapist 4.1 system will continue to the next image.

Firm Action

Siemens sent an "Urgent: Medical Device Correction/ Customer Safety Advisory Notice" letter to all affected customers starting June 3, 2011. The letter describes the product, problem, and instructions to prevent mistreatment. The letter states that Siemens is working on an updated version of RTT 4.1 that will fix the issues. Users will be notified as soon as the update for the system is available. The customers were instructed to include Customer Safety Advisory notice in their syngo RT Therapist 4.1 System Owner Manual chapter "Safety Advisory Letters" where it should remain.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - USA (nationwide) including the states of: DE, FL, KY, LA, MA, MI, MO, NE, NJ, OH, PA, TN, WI, UT, WV and Puerto Rico and the countries of: Australia, Belgium, Brazil, Canada, China, Egypt, France, Germany, Hungary, India, Ireland, Italy, Lebanon, Malaysia, Netherlands, New Zealand, Philippines, Poland, Republic of Korea, Russian Fed, Saudi Arabia, South Africa, Norway, Spain, Turkey and United Kingdom.