Software: Syngo.via RT Image Suite software version VA30_SP02 with license T_RT_IMAGE_SUITE_PAT_MARK Intended Use: used as an advanced application on a SOM10 scanner SMN: 10496180
FDA Recall Z-3005-2020 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeMUJ
Date Initiated2020-08-19
Date Terminated2023-04-19
Product Quantity64 systems US
Reason for Recall
Calibration offset between the central CT image plane and the external RT lasers has been set automatically to the factory default of 50.00 cm, if the misadjustment of the offset value is not recognized during the subsequent treatment planning process or remains undetected during the position verification at the treatment system (e.g. linear accelerator), this can potentially result in an incorrect dose or dose applied to the wrong location
Firm Action
Siemens Medical Solutions USA, Inc. issued Customer Safety Advisory Letter (CSAN) SY051/20/S on August 19, 2020 via email/hand delivery by Siemens Customer Service Engineer,. The letter states reason for recall, health risk and action to take: we highly recommend performing the laser calibration and the quality assurance procedure for the laser offset after every RT Image Suite software update.Please update the calibration offset in the MM RT Image Suite Configuration dialog accordingly . Details of this procedure are documented in Chapter 13.2 (Laser configuration) of your user manual.
If you have any questions or require technical support, please contact our service organization at 1-800-888-7436.
The technical solution distributed via SY040/20/S. Distribution of SY040/20/S (VB40B_HF01) will be distributed to all affected customers, free of charge. Distribution of SY040/20/S began on August 26, 2020. SY040/20/S will be installed with support from your local service organization
Root Cause
Software design
Distribution Pattern
US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.