AirLife Nebulizer Heater, Catalog Code 2M8021 The AirLife Nebulizer Heater is part of the heated nebulizer system intended to provide warm bland aerosol therapy to respiratory patients. The system is used in conjunction with a standard oxygen flowmeter, which by providing pressurized oxygen powers the nebulizer creating ultra-fine particles in the therapeutic aerosol. The heater warms the solution before it reaches the aerosol jet in the nebulizer and heats the dilution air before it is entrained into the aerosol stream. An aluminum heat exchanger on the nebulizer cap transfers the thermal energy to a nebulizer solution bottle warming the solution prior to aerosolizing.

FDA Recall Z-3016-2011 · Posted 2011-08-12

Recalling FirmCarefusion 211 Inc
Recall StatusTerminated
Product CodeCAF
Date Initiated2011-07-13
Date Terminated2012-04-30
Product Quantity22,670 units total

Reason for Recall

The recall was initiated because CareFusion's Investigation has confirmed the Nebulizer Heater can, over time, develop a failure mode that results in an internal electrical short. The short can result in smoke and sparks being emitted from the device. In the presence of oxygen enriched environments and a fuel source, this failure represents a potential risk of flame. CareFusion is voluntarily rec

Firm Action

Carefusion sent a URGENT PRODUCT RECALL letter dated July 15, 2011, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. Customers were instructed to examine their inventory for the affected product as indicated on the sample shown in the enclosed Attachment A-1 and remove any affected heaters present in their facility. Confirm the total number of affected units in their possession and enter the quantity and respective serial number on the enclosed Customer Response Card (Attachment A-2). Customers were instructed to contact CareFusion's Technical Support Department at 800-554-8933, (6:30 am to 5:00 pm PDST), to obtain an RMA (returned materials authorization) number for the return of the affected product. Upon receipt of their response, CareFusion will ship to their customers new replacement Nebulizer Heaters in exchange for the affected units. Customers were instructed to acknowledge receipt of this communication by completing the enclosed Customer Response Card and faxing or sending a copy by return mail to the address indicated on the Customer Response Card, (Attachment A-2).

Root Cause

Other

Distribution Pattern

Nationwide Distribution