RSP Standard HUMERAL SOCKET INSERT, 44MM, HXe-Plus, Part No. 509-00-044

FDA Recall Z-3021-2020 · Posted None

Recalling FirmEncore Medical, LP
Recall StatusOpen, Classified
Product CodePHX
Date Initiated2020-08-07
Product Quantity15 units

Reason for Recall

Product may be mislabeled with the size etched onto products differing from the size information shown on patient sticker labels. Risks associated with this issue may include the Surgeon implanting the incorrect size; significant physical impairment, discomfort and/or disruption of treatment of underlying condition; and the patient experiencing dislocation resulting in revision surgery.

Firm Action

Letters were sent to the firm's Field/Sales Agents by email beginning on August 7, 2020. The letters included instructions to take the following actions by August 21, 2020: 1. Pass the notice on to all those within your organization or to any organization where the potentially affected product has been transferred, 2.Contact Customer Service at 1-800-456- 8696 to place a replacement order and receive a Return Materials Authorization (RMA) number, and to be provided with a shipping label to return the affected product. 3. Return all affected devices using the RMA number, and 4. Report any adverse events related to this issue to Customer Service. Customers with questions should contact Christine Chesnutt at 512-834-6207 or [email protected].

Root Cause

Other

Distribution Pattern

Distributed in the U.S. to: VA, MN, IN, RI, PA, NY, WA, KS, OK, FL, GA and CA