ev3 TurboHawk PERPHERAL PLAQUE EXCISION SYSTEM FOR SURGICAL USE, LARGE VESSEL XTENDED TIP FOR CALCIUM, Model THS-LX-C for peripheral vascular use.

FDA Recall Z-3022-2018 · Posted None

Recall StatusTerminated
Product CodeMCW
Date Initiated2018-08-10
Date Terminated2020-08-04
Product Quantity67 units

Reason for Recall

The Distal Flush Tool (DFT) accessories that were packaged in this lot number are not compatible with the TurboHawk LX-C catheter. This may result in the inability to remove excised material from the tip of the catheter.

Firm Action

Medtronic began notifying the Consignees on 10-Aug-2018. Consignees were asked to quarantine and return affected product.

Root Cause

Process control

Distribution Pattern

Distribution to US states of KS, CA, FL, GA, AZ, MI, LA, TX, NE, KY, and China