Edwards Lifesciences Optisite Arterial Perfusion Cannula, 22F X 24CM, resterilized.
FDA Recall Z-3034-2018 · Posted None
Advertisement
Recalling FirmSPS Sterilization, Inc
Recall StatusTerminated
Product CodeDQR
Date Initiated2018-03-07
Date Terminated2018-11-27
Product Quantity1 unit total
Reason for Recall
The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
Firm Action
Customers were notified via letter on about 03/02/2018. Instructions included to review inventory for affected devices, arrange for their return to SPS Sterilization, and complete and return the Buisness Response Card.
A follow up letter was sent to customer on about 03/07/2018 and included a description of the risk and hazard involved. Instructions remained the same.
Root Cause
Labeling False and Misleading
Distribution Pattern
US distribution to AZ, CA, FL, OH, PA, TX, VA, WV
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.