Terminated

EMBLEM S-ICD, Model A209, Subcutaneous Implantable Cardioverter Defibrillator, RX. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing.

FDA Recall Z-3037-2017 · Posted 2017-08-16

Recall StatusTerminated
Product CodeLWS
Date Initiated2017-06-29
Date Terminated2019-02-21
Product QuantityApproximately 16,750 devices

Reason for Recall

The device can deliver an atypical amount of energy due to memory corruption inside the device.

Firm Action

The firm issued notifications dated June, 2017, beginning 6/29/2017 to physicians (implanting and patient follow-up physicians) via hand delivery by their sales representatives. The firm estimated approximately 30% of the U.S. physicians would be receiving the notification via hand delivery. Hand delivery by affiliates in foreign countries to customers began approximately 6/30/2017. Overnight mail letters were issued to the U.S. and foreign physicians starting on/about 7/7/2017 who did not receive hand- delivered notifications.

Root Cause

Software design

Distribution Pattern

Worldwide - US Nationwide distribution, including Puerto Rico, U.S. Virgin Island, and Guam, There was also worldwide foreign distribution, including Canada.