Terminated

ToxCup Drug Screen Cup (AMT Item Number/ AMT Description): HT29A/"ToxCup Drug Screen Cup COC300/MET500/OPI300/THC50/AMP1000/PCP25/OXY100 Adulteration Screen for CR - NI - PH - BL - S.G.", DT14/ "ToxCup Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; PCP25/PPX300", HT15/ "ToxCup Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000", HT15A/ ToxCup Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000 Adulteration Screen for CR - NI - PH - BL - S.G.", PT15/ "ToxCup Drug Screen Cup AMP1000/COC300/MET500/OPI300/THC50" DT13A/ "ToxCup Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100 w/ Cr, Ni, pH, BI, S.G.", DT14A "ToxCup Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; PCP25/PPX300 w/ Cr, Ni, pH, Bl, S.G. Finished Devices", DT12A/"ToxCup Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; MDMA500/OXY100; PCP25/PPX300 w/ Cr, Ni, pH, BI, S.G.", PT12A/ "ToxCup Drug Screen Cup COC300/MET500/OPI2000/PCP25/THC50 Adulteration Screen for CR - NI - PH - BL - S.G.", PT11A/ "ToxCup Drug Screen Cup AMP1000/COC300/OPI2000/PCP25/THC50 Adulteration Screen for CR - NI - PH - BL - S.G.", HT30/ "ToxCup Drug Screen Cup AMP1000/COC300/MET1000/PCP25/THC50/ OPI2000"

FDA Recall Z-3045-2020 · Posted None

Recalling FirmAmeditech Inc
Recall StatusTerminated
Product CodeDIO
Date Initiated2020-08-18
Date Terminated2024-09-30
Product Quantity189,175 units

Reason for Recall

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Firm Action

A Field Corrective Action Notice letter was sent to affected consignees. The Customer notifications commenced on 19-Aug-2020. The consignees are requested to return all non-expired product in their possession to the recalling firm. Method of Notification, 1st contact: Written field corrective action notification letters will be sent via certified mail with delivery tracking to each consignee along with a pre-paid return label to return any product remaining in stock. 2nd & 3rd contact: Telephone, follow up email, or written mail communication will be used as necessary.

Root Cause

Process control

Distribution Pattern

US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain