1) NEURO, (kit) ACNE73-01; 2) NEURO, (kit) ACNE73-03; 3) CRANI, (kit) AKCR19-01; 4) ORTHO SPINE PACK, (kit) ARMC017; 5) BMHM NEURO PACK, (kit) BMHT016-05; 6) LUMBAR PACK, (kit) BMHT112-03; 7) NEURO LAMINECTOMY PACK, (kit) DGNE75; 8) NEURO LAMINECTOMY PACK, (kit) DGNE75-02; 9) DRAH NEURO LAMINECTOMY SNEDHDN,(kit) DRNE75; 10) DRAH NEURO LAMINECTOMY SNEDHDN, (kit) DRNE75-01; 11) CRANIOSYNOSTOSIS CUSTOM SNE23C, (kit) DUCR36-01; 12) CRANIOSYNOSTOSIS CUSTOM SNE23C, (kit) DUCR36-02; 13) NEURO CRANIOTOMY PK SNE23C1116, (kit) DUNC42-01; 14) PEDS NEURO PACK SNE2301147, (kit) DUNE88-01; 15) NEURO SPINE PACK SNE23LM110, (kit) DUSP98-01; 16) NEURO SPINE PACK SNE23LM110, (kit) DUSP98-02; 17) CRANI PACK 712638, (kit) EFCR86-01; 18) SPINE PACK 712040, (kit) EFSP30-01; 19) SPINE PACK 712040, (kit) EFSP30-02; 20) CRANIOTOMY TRAY, (kit) EURO009-09; 21) CRANI PACK, (kit) IHCR36-01; 22) NEURO PACK, (kit) IHNE35-01; 23) CRANIOTOMY PACK, (kit) JACK410-11; 24) BASIC NEURO PACK, (kit) MAMC275-06; 25) BASIC NEURO PACK, (kit) MAMC275-07; 26) SPINE CRANI PACK, (kit) MCSP92-01; 27) CRANIOTOMY PACK, (kit) MECP05R; 28) SPINE TRAY GLOBALMICROSURG, (kit) MEDF217-02; 29) NEURO BASIC PACK, (kit) MENE06T; 30) CRANIAL PACK - 205992, (kit) MHCP75-01; 31) NEURO PACK, (kit) MMOK007-03; 32) NEURO PACK, (kit) MMTN012-04; 33) CRANIOTOMY PACK SNECGCRM11, (kit) MSMC108-01; 34) NEURO ANT CERV LAMI ADD A PACK, (kit) NAMC413-07; 35) STRL NEURO LUMB LAMI ADD A PK, (kit) NAMC433-05; 36) NEURO CRANI ADD A PACK, (kit) NAMC434-08; 37) NEURO PACK, (kit) NWHR010-03; 38) CRANIOTOMY PACK OSBORN 176518, (kit) OSCR20-01; 39) NEURO PACK, (kit) RDLD001-14; 40) CRANI PACK, (kit) RDLD008-14; 41) NEURO LUMBAR CERVICAL PACK,(kit) SAMM031-18; 42) CRANIOTOMY PACK, (kit) SELF111-02; 43) PELHAM ACDF, (kit) SGAC31-02; 44) PELHAM ACDF, (kit) SGAC31-03; 45) PELHAM MINA SPINE, (kit) SGSP11; 46) PELHAM MINA SPINE, (kit) SGSP11-03; 47) PELHAM SPINE, (kit) SGSP44-02; 48) PELHAM SPINE, (kit) SGSP44-03; 49) SPINE, (kit) SGSP91; 50) SPINE, (kit) SGSP91-02; 51) SPINE, (kit) SGSP91-03; 52) NEURO SPINAL TRAY SHANDS JACKS, (kit) SHND416-02; 53) DR. DELASOTTA NEURO NECK, (kit) SHOR003-15; 54) CRANI PACK, (kit) SLCN72-01; 55) CRANI PACK, (kit) SLCN72-02; 56) CRANI PACK, (kit) SLCN72-03; 57) CRANI PACK - 208400, (kit) UDCP57-01; 58) NEURO TRAY-LF, (kit) UMIA104-11; 59) CRANIOTOMY PACK, (kit) USMA005; 60) CERVICAL NEURO PACK PGYBK, (kit) VAML012-07; 61) LUMBAR NEURO PACK PGYBK, (kit) VAML028-06; 62) NASAL SINUS, (kit) VAPX009-02; 63) LUMBAR LAMINECTOMY TRAY, (kit) VMCC024-14; 64) VA LUMBAR LAMINECTOMY, (kit) VMED009-05; 65) ANTERIOR CERVICAL PACK, (kit) WEAC53-01; 66) CRANI PACK, (kit) WECR41-01; 67) LUMBAR LAMINECTOMY PACK,67) LUMBAR LAMINECTOMY PACK, Model/Item Number (kit) WELL73-01; 68) LUMBAR LAMINECTOMY, (kit) WKMC114-03; 69) CRANIOTOMY PK, (kit) WKMC125-03; 70) DISKECTOMY PACK, (kit) WRNM030-21.
FDA Recall Z-3045-2026 · Posted 2026-08-28
Advertisement
Recalling FirmAVID Medical, Inc.
Recall StatusOpen, Classified
Product CodeOJG
Date Initiated2026-05-19
Product Quantity16223 kits
Reason for Recall
The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
Firm Action
Owens & Minor issued an URGENT: MEDICAL DEVICE RECALL NOTIFICATION to its consignees on 5/19/2026 via email. The notice explained the issue, potential risk to health and requested the following:
"ACTIONS TO TAKE:
" Identify, segregate, and quarantine all impacted product.
" Add recall labels to affected Avid kits to ensure all users are aware of the actions to discard the impacted component. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling.
" Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have.
" Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness."
Root Cause
Material/Component Contamination
Distribution Pattern
US Nationwide distribution in the states of AZ, CA, CO, WA, WY, TX, MI, FL, MD, UT, NC, MS, PA, MO, IA, NE, OH, IL, SD, MN, GA.