Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Medline Convenience kits labeled as: 1) UROLOGY MINOR CDS , Pack Number CDS980371J ; 2) UROLOGY MINOR CDS , Pack Number CDS980371K ; 3) CIRCUMCISION TRAY 1.1 , Pack Number CIT6260A ; 4) CIRCUMCISION TRAY 1.3 CM BELL , Pack Number CIT7740; 5) CIRCUMCISION TRAY , Pack Number DYNDL1813; 6) CIRCUMCISION PACK , Pack Number DYNJ16826M ; 7) CIRCUMCISION PACK , Pack Number DYNJ16826N ; 8) CIRCUMCISION PACK , Pack Number DYNJ16826O ; 9) CIRCUMCISION PACK , Pack Number DYNJ38941D ; 10) CIRCUMCISION PACK , Pack Number DYNJ38941F ; 11) CIRCUMCISION SJHC-LF, Pack Number DYNJ48471D ; 12) CIRCUMCISION SJHC-LF, Pack Number DYNJ48471F ; 13) CRMC UNIVERSAL PACK-LF, Pack Number DYNJ61119C ; 14) D-CIRCUMCISION PACK , Pack Number DYNJ64432; 15) CIRCUMCISION PACK , Pack Number DYNJ64876A ; 16) CIRCUMCISION PACK , Pack Number DYNJ64876B ; 17) PACK,CIRCUMCISION ASC , Pack Number DYNJ906909A

FDA Recall Z-3047-2024 · Posted 2024-09-06

Recall StatusOpen, Classified
Product CodeOHG
Date Initiated2024-04-08

Reason for Recall

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

Firm Action

Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution.