First Check 4 Panel Drug Cup: COC/MET1000/OPI2000/THC): (THC/COC); AMT Item Number 990100B
FDA Recall Z-3049-2020 · Posted None
Advertisement
Recalling FirmAmeditech Inc
Recall StatusTerminated
Product CodeDKZ
Date Initiated2020-08-18
Date Terminated2024-09-30
Product Quantity537,120 units
Reason for Recall
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Firm Action
A Field Corrective Action Notice letter was sent to affected consignees. The Customer notifications commenced on 19-Aug-2020. The consignees are requested to return all non-expired product in their possession to the recalling firm.
Method of Notification, 1st contact: Written field corrective action notification letters will be sent via certified mail with delivery tracking to each consignee along with a pre-paid return label to return any product remaining in stock. 2nd & 3rd contact: Telephone, follow up email, or written mail communication will be used as necessary.
Root Cause
Process control
Distribution Pattern
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.