Baxter Novum IQ Syringe Pump, Model Number 40800BAXUS

FDA Recall Z-3065-2026 · Posted 2026-08-31

Recall StatusOpen, Classified
Product CodeFRN
Date Initiated2026-08-11
Product Quantity14601 units

Reason for Recall

A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.

Firm Action

On August 11, 2026, firm began sending customers Urgent Medical Device Correction delivered via email and USPS first class mail. Customers were advised to use the same compatible syringe model (brand/size) for all syringes required within a Multi-Syringe VOT workflow. Customers were provided with additional instructions for situations where the syringe does not support the intended flow rate and the user receives a prompt to confirm continuation of the infusion at the maximum or minimum supported flow rate for that syringe after a syringe change.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.