Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators DELUXE PRE OP KIT DYKS2018 GENERAL PREOP KIT DYKS1584A PRE OP KIT DYKS1216I PRE-OP KIT DYKM1398B

FDA Recall Z-3074-2026 · Posted 2026-08-31

Recall StatusOpen, Classified
Product CodeOJU
Date Initiated2026-07-20
Product Quantity222,653 total

Reason for Recall

Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.

Firm Action

Firm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution.