Terminated

VITEK 2 Compact 30, REF numbers: 27530 and 27530R (clinical), and 27630 and 17630R (industry) software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.

FDA Recall Z-3076-2017 · Posted None

Recalling FirmBiomerieux Inc
Recall StatusTerminated
Product CodeLON
Date Initiated2017-07-12
Date Terminated2020-12-16
Product Quantity9234 units

Reason for Recall

Customers have reported that some VITEK 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.

Firm Action

Consignees (Subsidiaries) of bioMrieux, impacted by the identified issue, were notified of the FSCA and corrective action (FSCA notification) electronically on 07/12/2017. The direct consignees (subsidiaries and distributors) have until 08/12/2017 to notify their respective impacted users. The directions to the user are as follows: Please implement the following actions at this time: " Confirm this letter has been distributed to, and reviewed by, all appropriate personnel within your organization. " Refrain from performing the VITEK 2 Systems Software 8.01 update. o A solution has been identified, and your local bioMrieux representative will contact you with further instruction in the coming weeks. " If you have updated VITEK 2 Systems Software to version 8.01 within the last seven (7) days, please contact your local bioMrieux representative. " Please store this letter with your bioMrieux instrument documentation. " Complete the Acknowledgement Form and return it to your local bioMrieux representative."

Root Cause

Other

Distribution Pattern

Worldwide