Fogarty Adherent Clot Catheter Model Numbers 1408010 and 140806 Product Usage: The Fogarty adherent clot catheters are designed to effectively remove clot material in the peripheral vasculature. The catheters are indicated for the removal of emboli and thrombi from native arteries or synthetic grafts. The device are designed to facilitate clots too resistant to be removed by an elastomeric balloon. The device features a spiral-shaped, latex-covered stainless steel cable that assumes a corkscrew shape when retracted, greatly expanding the surface areas to entrap fibrous material. The device is 80 cm in length and is available in 4F-6F diameters.
FDA Recall Z-3077-2017 · Posted None
Advertisement
Recalling FirmEdwards Lifesciences, LLC
Recall StatusTerminated
Product CodeDXE
Date Initiated2017-07-06
Date Terminated2018-10-17
Product Quantity200 units
Reason for Recall
The diameter of the Fogarty Catheter balloon, measures 6mm instead of 10mm.
Firm Action
Recall notification letters were sent to affected customers on 7/6/2017. The recall notice included a response form and instructed consignee to return any unused devices..
Root Cause
Packaging
Distribution Pattern
Worldwide Distribution - US Nationwide GB, NL, DE, IT, FR, BE, ES, NO, CZ, BG, SE, AT, CH, PL, RU, FI