LX3 Floor Stand for LuxOR Ophthalmic Microscope; Endure LuxOR Q-VUE Ophthalmic Microscope. LX3 Microscope Floor Stands are not stand-alone medical devices, but are used in combination with the LuxOR Ophthalmic Microscope and the LuxOR Ophthalmic Microscope with Q-VUE 3-D Assistant Visualization. These microscopes are used for low magnification visualization during cataract, retina, and cornea surgeries

FDA Recall Z-3081-2017 · Posted None

Recalling FirmAlcon Research, Ltd.
Recall StatusTerminated
Product CodeHRM
Date Initiated2017-05-25
Date Terminated2022-02-04
Product Quantity1,074 units

Reason for Recall

Multiple shipments or forceful impact to a caster on a LuxOr/LX3 microscope may stress the microscope base, which could contribute to a caster breaking or detaching from the base.

Firm Action

Alcon sent an Urgent Medical Device Correction letter dated May 25, 2017, to all affected customers. Demonstration units in inventory will be scheduled for retrofit of base and/or upgraded caster configuration through onsite service calls. Customers with questions or concerns were instructed to call 1-800-832-7827. For questions regarding this recall call 817-293-0450.

Root Cause

Packaging process control

Distribution Pattern

Worldwide Distribution - US (nationwide and Puerto Rico): Units were also distributed globally to the following countries: Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bosnia, Herzegovina, Bulgaria, Canada, Chile, Columbia, Costa Rica, Croatia, Curacao, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Ecuador, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hungary, India, Italy, Jamaica, Japan, Jordan, Kazakhstan, South Korea, Kuwait, Lebanon, Lithuania, Luxembourg, Malaysia, Mexico, Mongolia, Morocco, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, Ukraine, United Kingdom, and Uzbekistan