Maquet HL Console for Heart Lung Machine HL20 HL 20 4-pumps console base: P/N 917741 HL 20 5-pumps console base: P/N 917751 Intended for use as an extracorporeal circulation device for perfusion lasting not more than six hours. The HL console is equipped with a rack for accommodating the system's electronic modules.
FDA Recall Z-3084-2011 · Posted 2011-08-29
Advertisement
Recalling FirmMaquet Inc.
Recall StatusTerminated
Product CodeDTQ
Date Initiated2011-06-20
Date Terminated2012-10-16
Product Quantity299 units
Reason for Recall
Maquet HL 20 Twin Pump Modules may unintentionally increase speeds for a 1.5-2 second period of time before returning to its original speed.
Firm Action
Maquet sent a PRODUCT RECALL NOTIFICATION LETTER/return response forms, dated June 17, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product until a software update was performed by their trained technicians. Customers were also asked to forward the information contained in the notification letter to all users and to all appropriate Biomedical Staff.
Customers were instructed to acknowledge receipt of the notification letter by completing and faxing back the attached response form to 973-807-9210 whether or not they have any of the affected product.
For questions customers were instructed to contact their local MAQUET representative.
For questions regarding this recall call 973-709-7994.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.