smiths medical portex BLUselect, tracheostomy tube, labeled as follows: 1) BLUSELECT 6.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/060CZ; 2) BLUSELECT 7.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/070CZ; 3) BLUSELECT 7.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/075CZ; 4) BLUSELECT 8.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/080CZ; 5) BLUSELECT 8.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/085CZ; 6) BLUSELECT 9.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/090CZ; 7) BLUSELECT 10.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/100CZ; 8) BLUSELECT 6.0, CUFFED, FEN, 10/BX, Item Number 101/802/060CZ; 9) BLUSELECT 7.0, CUFFED, FEN, 10/BX, Item Number 101/802/070CZ; 10) BLUSELECT 7.5, CUFFED, FEN, 10/BX, Item Number 101/802/075CZ; 11) BLUSELECT 8.0, CUFFED, FEN, 10/BX, Item Number 101/802/080CZ; 12) BLUSELECT 8.5, CUFFED, FEN, 10/BX, Item Number 101/802/085CZ; 13) BLUSELECT 9.0, CUFFED, FEN, 10/BX, Item Number 101/802/090CZ; 14) BLUSELECT 10.0, CUFFED, FEN, 10/BX, Item Number 101/802/100CZ; 15) BLUSELECT 6.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/060CZ; 16) BLUSELECT 7.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/070CZ; 17) BLUSELECT 7.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/075; 18) BLUSELECT 7.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/075CZ; 19) BLUSELECT 8.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/080; 20) BLUSELECT 8.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/080CZ; 21) BLUSELECT 8.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/085CZ; 22) BLUSELECT 9.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/090CZ; 23) BLUSELECT 10.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/100CZ; 24) BLUSELECT 6.0, CUFFED, FEN, 1/EA, Item Number 101/812/060CZ; 25) BLUSELECT 7.0, CUFFED, FEN, 1/EA, Item Number 101/812/070CZ; 26) BLUSELECT 7.5, CUFFED, FEN, 1/EA, Item Number 101/812/075CZ; 27) BLUSELECT 8.0, CUFFED, FEN, 1/EA, Item Number 101/812/080CZ; 28) BLUSELECT 8.5, CUFFED, FEN, 1/EA, Item Number 101/812/085CZ; 29) BLUSELECT 9.0, CUFFED, FEN, 1/EA, Item Number 101/812/090CZ; 30) BLUSELECT 10.0, CUFFED, FEN, 1/EA, Item Number 101/812/100CZ; 31) BLUSELECT 6.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/060; 32) BLUSELECT 7.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/070; 33) BLUSELECT 7.5, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/075; 34) BLUSELECT 8.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/080; 35) BLUSELECT 8.5, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/085; 36) BLUSELECT 9.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/090; 37) BLUSELECT 10.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/100; 38) BLUSELECT 6.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/060; 39) BLUSELECT 7.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/070; 40) BLUSELECT 8.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/080; 41) BLUSELECT 9.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/090; 42) BLUSELECT 10.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/100; 43) BLUSELECT 6.0, SUCTIONAID, CUFFED, NON-FEN,10/BX, Item Number 101/860/060CZ; 44) BLUSELECT 7.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/070CZ; 45) BLUSELECT 7.5, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/075CZ; 46) BLUSELECT 8.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/080CZ; 47) BLUSELECT 8.5, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/085CZ; 48) BLUSELECT 9.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/090CZ; 49) BLUSELECT 10.0, SUCTIONAID, CUFFED, NON-FEN,10/BX, Item Number 101/860/100CZ; 50) BLUSELECT 6.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/060CZ; 51) BLUSELECT 7.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/070CZ; 52) BLUSELECT 7.5, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/075CZ; 53) BLUSELECT 8.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/0

FDA Recall Z-3086-2024 · Posted 2024-09-11

Recall StatusOpen, Classified
Product CodeBTO
Date Initiated2024-06-07
Product Quantity799732 units

Reason for Recall

There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.

Firm Action

Smiths Medical issued an URGENT MEDICAL DEVCIE NOTIFICATION to its consignees on 06/07/2024 via traceable mailing. The notice explained the issue with the device, potential risk to the patient, and requested the affected devices be discarded. Distributor were directed to forward the notice to their customers. For questions: Customer service: [email protected]; 800-258-5361) Complaints: [email protected]; 866-216-8806

Root Cause

Process control

Distribution Pattern

Worldwide