smiths medical portex BLUselect, tracheostomy tube, labeled as follows: 1) BLUSELECT 6.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/060CZ; 2) BLUSELECT 7.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/070CZ; 3) BLUSELECT 7.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/075CZ; 4) BLUSELECT 8.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/080CZ; 5) BLUSELECT 8.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/085CZ; 6) BLUSELECT 9.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/090CZ; 7) BLUSELECT 10.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/100CZ; 8) BLUSELECT 6.0, CUFFED, FEN, 10/BX, Item Number 101/802/060CZ; 9) BLUSELECT 7.0, CUFFED, FEN, 10/BX, Item Number 101/802/070CZ; 10) BLUSELECT 7.5, CUFFED, FEN, 10/BX, Item Number 101/802/075CZ; 11) BLUSELECT 8.0, CUFFED, FEN, 10/BX, Item Number 101/802/080CZ; 12) BLUSELECT 8.5, CUFFED, FEN, 10/BX, Item Number 101/802/085CZ; 13) BLUSELECT 9.0, CUFFED, FEN, 10/BX, Item Number 101/802/090CZ; 14) BLUSELECT 10.0, CUFFED, FEN, 10/BX, Item Number 101/802/100CZ; 15) BLUSELECT 6.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/060CZ; 16) BLUSELECT 7.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/070CZ; 17) BLUSELECT 7.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/075; 18) BLUSELECT 7.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/075CZ; 19) BLUSELECT 8.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/080; 20) BLUSELECT 8.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/080CZ; 21) BLUSELECT 8.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/085CZ; 22) BLUSELECT 9.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/090CZ; 23) BLUSELECT 10.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/100CZ; 24) BLUSELECT 6.0, CUFFED, FEN, 1/EA, Item Number 101/812/060CZ; 25) BLUSELECT 7.0, CUFFED, FEN, 1/EA, Item Number 101/812/070CZ; 26) BLUSELECT 7.5, CUFFED, FEN, 1/EA, Item Number 101/812/075CZ; 27) BLUSELECT 8.0, CUFFED, FEN, 1/EA, Item Number 101/812/080CZ; 28) BLUSELECT 8.5, CUFFED, FEN, 1/EA, Item Number 101/812/085CZ; 29) BLUSELECT 9.0, CUFFED, FEN, 1/EA, Item Number 101/812/090CZ; 30) BLUSELECT 10.0, CUFFED, FEN, 1/EA, Item Number 101/812/100CZ; 31) BLUSELECT 6.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/060; 32) BLUSELECT 7.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/070; 33) BLUSELECT 7.5, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/075; 34) BLUSELECT 8.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/080; 35) BLUSELECT 8.5, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/085; 36) BLUSELECT 9.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/090; 37) BLUSELECT 10.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/100; 38) BLUSELECT 6.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/060; 39) BLUSELECT 7.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/070; 40) BLUSELECT 8.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/080; 41) BLUSELECT 9.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/090; 42) BLUSELECT 10.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/100; 43) BLUSELECT 6.0, SUCTIONAID, CUFFED, NON-FEN,10/BX, Item Number 101/860/060CZ; 44) BLUSELECT 7.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/070CZ; 45) BLUSELECT 7.5, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/075CZ; 46) BLUSELECT 8.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/080CZ; 47) BLUSELECT 8.5, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/085CZ; 48) BLUSELECT 9.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/090CZ; 49) BLUSELECT 10.0, SUCTIONAID, CUFFED, NON-FEN,10/BX, Item Number 101/860/100CZ; 50) BLUSELECT 6.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/060CZ; 51) BLUSELECT 7.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/070CZ; 52) BLUSELECT 7.5, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/075CZ; 53) BLUSELECT 8.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/0
FDA Recall Z-3086-2024 · Posted 2024-09-11
Advertisement
Recalling FirmSmiths Medical ASD Inc.
Recall StatusOpen, Classified
Product CodeBTO
Date Initiated2024-06-07
Product Quantity799732 units
Reason for Recall
There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.
Firm Action
Smiths Medical issued an URGENT MEDICAL DEVCIE NOTIFICATION to its consignees on 06/07/2024 via traceable mailing. The notice explained the issue with the device, potential risk to the patient, and requested the affected devices be discarded. Distributor were directed to forward the notice to their customers.
For questions: Customer service: [email protected]; 800-258-5361)
Complaints: [email protected]; 866-216-8806
Root Cause
Process control
Distribution Pattern
Worldwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K173912BLUperc Percutaneous Dilation Tracheostomy Kit with BLUselect tube, BLUperc Percutaneous Dilation Tracheostomy Kit with BLUselect Suctionaid tube, BLUgriggs Percutaneous Dilation Tracheostomy Kit with BLUselect tube, BLUgriggs Percutaneous Dilation Tracheostomy Kit with BLUselect Suctionaid tubeSmiths Medical Asd, Inc. · 2018-05-08K173384BLUselect, non-fenestrated, BLUselect, non-fenestrated, cont, BLUselect, fenestrated, BLUselect Suctionaid, BLUpere Dilation Trachestomy Kit, without tubeSmiths Medical Ads, Inc. · 2018-04-10
Official Source