5008X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8

FDA Recall Z-3087-2026 · Posted 2026-09-02

Recall StatusOpen, Classified
Product CodeKOC
Date Initiated2026-08-10
Product Quantity6,624 units

Reason for Recall

Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.

Firm Action

On 8/10/2026, the firm began notifying consignees via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to 1. Identify any affected product on hand. 2. IMMEDIATELY segregate and discontinue use of affected lot. 3. Contact the call center at 1-888-943-4901 and reference Event 8514, between 8:00 AM 4:30 PM Eastern daily, for return of affected lots and replacement of your stock. 4, Complete and return the included reply form.

Root Cause

Component change control

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.