Open, Classified

Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus.

FDA Recall Z-3090-2026 · Posted 2026-09-11

Recalling FirmGiven Imaging Ltd.
Recall StatusOpen, Classified
Product CodeFFT
Date Initiated2026-08-06
Product Quantity127,416

Reason for Recall

Capsule may fail to attach to the esophagus or detach from the delivery device. Risks include patient aspiration/inhalation, perforation of esophagus, obstruction of the airway, hemorrhage/blood loss/bleeding, laceration of the esophagus, a delay in diagnosis, and foreign bodies remaining in the patient.

Firm Action

On August 10, 2026, the firm notified consignees via "URGENT: MEDICAL DEVICE RECALL" letters. Please note this is a separate recall from that initiated in June 2025, which involved the same issue, but a different underlying cause. Customers were instructed to do the following: 1. Immediately identify and quarantine all unused and affected devices. 2. Return all product to Medtronic for replacement as described in the letter. 3. Complete and return the included customer response form to the firm via email: [email protected], even if you do not have unused affected inventory. 4. Notify all those who need to be aware within your organization, or those to whom affected product was transferred/distributed. 5. Maintain a copy of the recall letter for your records. If you have any questions regarding this communication, please contact your local Medtronic Representative or call 720-774-3454.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution. US nationwide including Puerto Rico; Argentina, Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Czech Republic, Dominican Republic, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Kazakhstan, Kenya, Luxembourg, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, and United Kingdom.

Official Source

View FDA Recall →