On-Board Imager (OBI) 1.3, 1.4, 1.5; and Trilogy Mx, TrueBeam 1.0, Offline Review 1.0 - 2.0; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields relation to anatomical and/or fiducial landmarks. Trilogy Mx is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
FDA Recall Z-3101-2011 · Posted 2011-08-31
Advertisement
Recalling FirmVarian Medical Systems, Inc. Oncology Systems
Recall StatusTerminated
Product CodeIYE
Date Initiated2011-06-21
Date Terminated2012-03-15
Product Quantity2268
Reason for Recall
Varian has identified a software anomaly in the 3D Match environment, regarding a slight mismatch between CT image and Structures in On-Board Imager w/s.
Firm Action
Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated June 21, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers.
***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
The letter recommends several user actions. The firm is developing a correction for the issue and will contact customers when the correction is available.
Questions regarding this notice may be addressed by contacting the Varian Oncology Help Desk at 1-888-827-4265.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution