Open, Classified

Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ASLMC CRANIOTOMY , Catalog Number: PNCGCNAUM; 2) KIT, ORP MAJOR SPINE , Catalog Number: PNHDFSVCI; 3) KIT,ORP NEURO TRAUMA , Catalog Number: PNHDOTVCQ; 4) KIT, CRANIOTOMY PRESBYTERIAN , Catalog Number: PNV20101L; 5) KIT, LARGE SPINE 143487 , Catalog Number: PNV20101L; 6) FUSION ACCESSORY PACK BRO , Catalog Number: SNECGLFABB; 7) LH GENERAL NEURO SPINE PACK , Catalog Number: SNEOCGSLHP; 8) LH GENERAL NEURO SPINE PACK , Catalog Number: SNEOCGSL20

FDA Recall Z-3104-2026 · Posted 2026-09-03

Recall StatusOpen, Classified
Product CodeOJG
Date Initiated2026-07-29
Product Quantity645 units

Reason for Recall

The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.

Firm Action

An URGENT: PRODUCT CORRECTION notification letter dated 7/29/26 was sent to customers. If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: [email protected] For component credit or replacement, please contact the following: " Hospitals Cardinal Customer Service at 866.551.0533 " Distributors Distributor Management at 800.635.6021 " Federal Government facilities Government Customer Service at 800.444.1166 If you are a Distributor and have further distributed affected kits please notify your customers of this recall. We sincerely apologize for any inconvenience this notice may have caused you and your staff. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.

Root Cause

Process change control

Distribution Pattern

US Nationwide distribution.