Depuy Mitek Ratchet Handle with Quick Connect Product code: 219970. Instrument used with anchor inserters for insertion of DePuy Mitek Interference Screws (orthopedic).

FDA Recall Z-3106-2011 · Posted 2011-08-31

Recall StatusTerminated
Product CodeLXH
Date Initiated2011-06-27
Date Terminated2012-01-19
Product Quantity75 units

Reason for Recall

Quick Disconnect Ratchet Handle may slip up the driver shaft or disengage leaving the shaft in the screw head.

Firm Action

Depuy Mitek issued Original Customer Notification Letter on 6/27/11 and an Updated letter to clarify the lot number information sent 6/29/11. Hospitals requested to immediately check all inventories for affected product and return for repalcement product. Also notify surgeons aware of this potential issue and potential risk of a delay in surgery. questions or concerns with regard to this program, please contact your local DePuy Mitek Representative or DePuy Mitek Customer Service directly at 1-800-382-4682, PROMPT 3.

Root Cause

Process change control

Distribution Pattern

Nationwide distribution.