Skintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only. Used with automated external defibrillators

FDA Recall Z-3109-2017 · Posted 2017-09-12

Recall StatusTerminated
Product CodeMKJ
Date Initiated2017-08-08
Date Terminated2020-06-09
Product Quantity8,249 units total

Reason for Recall

Certain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.

Firm Action

On approximately 08/08/2017, customers were notified via email of the recall. Instructions included to ensure all users and anyone concerned are aware of the recall, coordinate the replacement of affected product, once replacement electrodes are received document destruction of affected product on-site, and notify customers if the product had been further distributed.

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution. US distribution in the states of : FL & TX. Austria, China, France, Germany, Greece, Hungary, India, Italy, Panama, Philippines, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, United Arab Emirates, United Kingdom.