Halyard IV Start Kit Non-Sterile REF: KRIV26-02
FDA Recall Z-3109-2026 · Posted 2026-09-04
Advertisement
Recalling FirmAVID Medical, Inc.
Recall StatusOpen, Classified
Product CodeLRO
Date Initiated2026-07-14
Product Quantity98 Kits
Reason for Recall
Convenience kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which lacks sterility assurance due to to wrinkles in the paper lidding which may breach the seal area.
Firm Action
On July 14, 2026, Avid Medical Inc. a subsidiary of Owens Minor issued an Urgent Medical Device Recall to affected consignees via email. I Avid Medical Inc. requested consignees to take the following actions:
1. Identify, segregate, and quarantine all impacted product identified.
1a. Discontinue use of the BD ChloraPrep" Clear - 1 mL and FREPP" Clear 1.5 mL Applicators.
1b. Add warning labels to affected AVID kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical
product information found on existing product labeling.
2. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have.
3. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness.
4. Complete the attached Recall Response Form and return by July 31, 2026.
Root Cause
Packaging process control
Distribution Pattern
US Nationwide distribution in the state of NE.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.