Terminated

Toshiba Medical Radrex - i Digital Radiography X-ray system Radrex-i is intended for use in conjunction with the ceiling suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and trans mission/ output of images to external devices.

FDA Recall Z-3111-2017 · Posted None

Recall StatusTerminated
Product CodeMQB
Date Initiated2017-03-23
Date Terminated2020-10-23
Product Quantity140

Reason for Recall

It was discovered during a procedure that when the operator made an exposure on the wireless x-ray detector and the image data was sent to the digital radiography system, and error message was displayed " System Error (2063)" which required a reboot and loss of the image.

Firm Action

Toshiba America Medical Systems Inc. planned action to bring defect into compliance: 1.You will issue a Customer Notification Letter to make customers aware of the hazards, what measures will be taken to remedy the defect at no cost to the user, and how to safely use the equipment before corrections can be made. 2.You will install modified firmware on the systems to correct the defect and prevent recurrence. 3.You will provide progress reports on the effectiveness of the plan. If you have any questions, please feel free to contact your local Toshiba representative at (800) 421-1968.

Root Cause

Software design

Distribution Pattern

USA (nationwide) Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →