Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and Sekisui Diagnostics P.E.I. Inc., Canada. Used in laboratory testing.
FDA Recall Z-3118-2017 · Posted None
Advertisement
Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeLDP
Date Initiated2017-07-20
Date Terminated2018-01-04
Product Quantity320 units
Reason for Recall
Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification.
Firm Action
The recalling firm issued letters dated 7/17/2017 via regular mail and email on 7/20/2017 to all customers receiving the affected product.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Distribution US nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.