HeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes: 0506016712067 350P US Device, US English (American Airlines) - 2 Min CPR UDI-DI code: 0506016712207 350P US Device, US English with Gateway - 2 Min CPR UDI-DI code: 0506016712804 HeartSine SAM 360P 360P US Device, US English - 2 Min CPR UDI-DI code: 0506016712068 HeartSine SAM 450P SAM 450P US Device, US English (2 min CPR) UDI-DI code: 0506016712069 The HeartSine samaritan PAD (Public Access Defibrillator) Models SAM 350P, SAM 360P and SAM 450P are indicated for use on victims of cardiac arrest who are exhibiting the following signs: " Unconscious " Not breathing " Without circulation (without a pulse) The devices are indicated for use on patients greater than 8 years old or over 55 lbs. (25 kg) when used with the adult Pad-Pak (Pad-Pak-01 or Pad-Pak-07). They are indicated for use on children between 1 and 8 years of age or up to 55 lbs. (25 kg) when used with the Pediatric-Pad (Pad-Pak-02).

FDA Recall Z-3120-2026 · Posted 2026-09-04

Recall StatusOpen, Classified
Product CodeMKJ
Date Initiated2026-07-27
Product Quantity11,912

Reason for Recall

Due to a damaged (over stretched) membrane tail that causes an intermittent fault.

Firm Action

On 07/28/2026, several versions of an "Urgent: Medical device recall" letter that outlines multiple issues: 1) Stryker identified an issue with the membrane tail, an internal component, which could cause an intermittent fault. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. The device becomes inoperable after the failure occurs. 2) Stryker has also identified that a limited number of HeartSine samaritan PAD devices were distributed without the required maintenance insert. This insert contains periodic maintenance instructions implemented as an additional risk control measure. While most devices contain the maintenance insert, this voluntary correction is being initiated to ensure all end users have access to the supplemental maintenance information. This issue was identified through a customer complaint. For the membrane tail issue (issue 1), Customers are instructed to:

Root Cause

Process control

Distribution Pattern

Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME MI, MN, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SD,TN, TX, UT, VA, WA, WI, WV and WY. The countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Hong Kong, Indonesia, ISRAEL, JAPAN, Malaysia, Netherlands, Philippines, POLAND, SINGAPORE, South Korea, Taiwan, Thailand and UNITED KINGDOM.