Open, Classified

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) LAMINECTOMY CDS, Pack Number CDS981510L; b) ORTHO TRAUMA-LF, Pack Number DYNJ900720D; c) ORTHO TRAUMA-LF, Pack Number DYNJ900720F; d) KNEE ARTHROSCOPY, Pack Number DYNJ908090C; e) LAMINECTOMY, Pack Number DYNJ908091B; f) PODIATRY, Pack Number DYNJ908092C; g) ANTERIOR TOTAL HIP, Pack Number DYNJ908095B; h) CRANI, Pack Number DYNJ908097A; i) HIP PINNING, Pack Number DYNJ908102B; j) (76) LOWER EXTREMITY, Pack Number DYNJ908103B; k) SHOULDER ARTHROSCOPY, Pack Number DYNJ908110B; m) HAND-LF, Pack Number DYNJ908425; n) TOTAL JOINT, Pack Number DYNJ908660A; o) MERCY TIFFIN HAND, Pack Number DYNJ909734; p) MERCY TIFFIN TOTAL KNEE, Pack Number DYNJ909738; q) MTHZ KNEE ARTHROSCOPY, Pack Number DYNJ909740; r) ACL-WHSC, Pack Number DYNJ910074A; s) EXTREMITY-WHSC, Pack Number DYNJ910077; t) EXTREMITY-WHSC, Pack Number DYNJ910077A; u) KNEE ARTHROSCOPY-WHSC, Pack Number DYNJ910080A; v) MICRODISC, Pack Number DYNJ910370; w) MTHZ TOTAL HIP, Pack Number DYNJC9311L

FDA Recall Z-3121-2024 · Posted 2024-09-11

Recall StatusOpen, Classified
Product CodeOJH
Date Initiated2024-06-27
Product Quantity1076 units

Reason for Recall

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Firm Action

Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed. For questions: contact the Recall Department at 866-359-1704 or [email protected].

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Panama.