Cell Marque, MSH6 (44 Mab.), 7 mL, REF PA0804, IVD, The antibody is intended for in vitro diagnostic (IVD) use. MSH6 is a mismatch repair gene which is deficient in a high proportion of patients with microsatellite instability (MSI-H). This finding is associated with the autosomal dominant condition known as Hereditary Non-Polyposis Colon Cancer (HNPCC).

FDA Recall Z-3123-2017 · Posted None

Recall StatusTerminated
Product CodeNJT
Date Initiated2017-07-14
Date Terminated2024-08-02
Product Quantity13

Reason for Recall

A delay in transit may have contributed to nonconformity as the product was stored outside of defined shipping and handling temperature requirements.

Firm Action

Advance Cancer Diagnostics sent an Urgent Medical Device Recall letter dated July 12, 2017, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Customers were instructed to identify any affected products, discard them, and complete and return the Medical Device Acknowledgement Form by fax to 1-847-236-3000. For questions regarding this recall call 815-678-2000.

Root Cause

Storage

Distribution Pattern

Nationwide Distribution to AR, AZ, CA, CT, FL, GA, IL, IN, KY, LA, MI, MN, NC, NH, NV, NY, OR, PA, SC, TN, TX, VA, and WI.