Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV
FDA Recall Z-3143-2026 · Posted 2026-09-11
Advertisement
Recalling FirmGE Medical Systems Information Technologies Inc
Recall StatusOpen, Classified
Product CodeDQK
Date Initiated2026-07-30
Product Quantity29,570 (23,632 US; 5,938 OUS)
Reason for Recall
Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
Firm Action
On July 30, 2026, the firm sent Urgent Medical Device Correction letters to
customers with affected units. This letter informs the customer of the issue and provides
actions to be taken by the customer/user for continued use of the device. GE HealthCare will provide a User Manual Addendum that provides guidance regarding the prolonged startup condition.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide - US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.