Flexima Regular Kit Biliary Catheter System Kit, Material Numbers REF M001271650; bile drainage catheter with locking pigtail
FDA Recall Z-3148-2024 · Posted 2024-09-13
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusOpen, Classified
Product CodeFFA
Date Initiated2024-07-18
Product Quantity32 units
Reason for Recall
There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.
Firm Action
Boston Scientific issued am Urgent Medical Device Removal notice to it consignees on 07/18/2024 via overnight mail. The notice explained the issue, potential hazard, and requested the return of the devices. Consignees were directed to post the notice in a visible location near the affected products to ensure visibility and to share the notice with all those to whom the device was transferred. Distributors were direct to notify their customers.
Root Cause
Packaging process control
Distribution Pattern
US, Colombia, Philippines, Russia
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.