VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the urinary tract.

FDA Recall Z-3151-2024 · Posted 2024-09-13

Recall StatusOpen, Classified
Product CodeFFA
Date Initiated2024-07-18
Product Quantity5 units

Reason for Recall

There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.

Firm Action

Boston Scientific issued am Urgent Medical Device Removal notice to it consignees on 07/18/2024 via overnight mail. The notice explained the issue, potential hazard, and requested the return of the devices. Consignees were directed to post the notice in a visible location near the affected products to ensure visibility and to share the notice with all those to whom the device was transferred. Distributors were direct to notify their customers.

Root Cause

Packaging process control

Distribution Pattern

US, Colombia, Philippines, Russia