Haylard UNIV IMAGING BIOPSY TRUCUSTOM; Model/Item Number (Kit): VARV042-03;
FDA Recall Z-3154-2026 · Posted 2026-09-16
Advertisement
Recalling FirmAVID Medical, Inc.
Recall StatusOpen, Classified
Product CodeOES
Date Initiated2026-08-10
Product Quantity340 kits
Reason for Recall
Potential for product quality and sterility issues regarding kits containing Sodium Chloride ampules subject to recall by Spectra Medical.
Firm Action
On August 10, 2026 URGENT: MEDICAL DEVICE RECALL letters were sent to customers.
Actions to be taken:
1. Identify, Segregate, and Quarantine all impacted kits listed.
2. Add warning labels to affected kits to ensure all users are aware of the actions to take prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. If pre-printed labels are needed, please contact your sales representative.
3. REMOVE AND DISCARD the Sodium Chloride, 0.9%, USP 10 mL Flush ampules prior to use.
4. Customers (end-users): If you have received product from a distributor, please contact the distributor or your sales representative with any questions you may have.
5. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness.
6. Complete the attached Recall Response Form and return.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Domestic: IL, NC, VA, WY;
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.