Mazor X. Model Number REF TPL0059; stereotaxic orthopedic instrument
FDA Recall Z-3156-2024 · Posted 2024-09-16
Advertisement
Recalling FirmMazor Robotics Ltd
Recall StatusOpen, Classified
Product CodeOLO
Date Initiated2024-07-17
Product Quantity94 devices
Reason for Recall
Software anomalies which can result in two separate issues: A lagging screen display during a navigated procedure; or the potential for a mismatch between the navigated screw and the screw information presented on the screen.
Firm Action
Medtronic/Mazor Robotics, Ltd. issued an URGENT: Medical Device Software Correction to its consignees on 07/17/2024 via 2-day tracked mailing. The notice explained the issue, potential risk, and provided a workaround until the software update is installed. Medtronic will schedule a time to perform a software update that will resolve these
anomalies. Medtronic field service engineers will update all Mazor X systems from version 5.1.1 to version 5.1.2 or later.
Questions: contact Medtronic Technical Services at 1-888-826-5603 or [email protected] .
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution in PR.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.