Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment
FDA Recall Z-3162-2024 · Posted 2024-09-17
Advertisement
Recalling FirmBtt Health Gmbh
Recall StatusOpen, Classified
Product CodeBXB
Date Initiated2024-08-01
Product Quantity62 units
Reason for Recall
Some power supply units in Austria and Germany had a voltage drop in the millisecond range, which led to a cancellation of the treatment and subsequent restart of the medical device. The treatment is only restarted by a step-off/step-on (step-on for activation).
Firm Action
Btt notified its distributors on 08/01/2024 via email. The notice explained the problem with the device, impact the the user, and were directed to notify their customers. The power supplies will be replaced.
Root Cause
Component design/selection
Distribution Pattern
US Nationwide distribution in the states of Florida.