DIAMOND POINT, BOTH ENDS K-WIRE, REF 128042, metallic bone fixation fastener
FDA Recall Z-3171-2024 · Posted 2024-09-19
Advertisement
Recalling FirmSmith & Nephew, Inc.
Recall StatusOpen, Classified
Product CodeHTY
Date Initiated2024-07-09
Product Quantity24 units
Reason for Recall
Smith & Nephew, Inc., has initiated a field action to voluntarily remove two batches of K-Wires due to a packaging error. A complaint was received indicating that a package contained K-WIRE .062X9 2PT DM wires instead of K-WIRE .045X9 2PT DM as described on the product label.
Firm Action
Smith & Nephew issued and Urgent Medical Device Recall Notice to its consignees on 07/09/2024 via overnight mail. The notice explained the problem, potential risk, and requested the return of the product. Sales representatives, District Offices, and/or Distributors were directed to notify their customers.
For questions or concerns regarding this recall please contact [email protected].
Root Cause
Packaging change control
Distribution Pattern
US, Canada, Norway
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.