Kinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide), REF: 5831-9006
FDA Recall Z-3178-2026 · Posted 2026-09-16
Advertisement
Recalling Firmrestor3d Inc.
Recall StatusOpen, Classified
Product CodeOYK
Date Initiated2026-07-27
Product Quantity2
Reason for Recall
Ankle implantation systems were packaged with incorrect patient-specific cutting guides, which may result in: delayed surgery, the need to use standard instrumentation, reduced fixation or poor alignment of implants, implant loosening/migration, or need for revision surgery.
Firm Action
On 7/28/2026, recall notices were emailed to customers who were asked to do the following:
1) Segregate and return affected devices.
2) Complete and return the response form.
If additional questions, contact firm at 984-201-2310
Root Cause
Labeling Change Control
Distribution Pattern
US Nationwide distribution in the states of FL, CO.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.