Kinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide), REF: 5831-9006

FDA Recall Z-3178-2026 · Posted 2026-09-16

Recalling Firmrestor3d Inc.
Recall StatusOpen, Classified
Product CodeOYK
Date Initiated2026-07-27
Product Quantity2

Reason for Recall

Ankle implantation systems were packaged with incorrect patient-specific cutting guides, which may result in: delayed surgery, the need to use standard instrumentation, reduced fixation or poor alignment of implants, implant loosening/migration, or need for revision surgery.

Firm Action

On 7/28/2026, recall notices were emailed to customers who were asked to do the following: 1) Segregate and return affected devices. 2) Complete and return the response form. If additional questions, contact firm at 984-201-2310

Root Cause

Labeling Change Control

Distribution Pattern

US Nationwide distribution in the states of FL, CO.