VITEK(R) 2 AST-P640, For susceptibility testing of Staphylococcus spp. Enterococcus spp., ad S agalactiae against specified antimicrobials, REF 418579 Product Usage: VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.
FDA Recall Z-3179-2018 · Posted None
Advertisement
Recalling FirmbioMerieux, Inc.
Recall StatusTerminated
Product CodeLCD
Date Initiated2018-08-13
Date Terminated2020-11-02
Product Quantity2538 cartons
Reason for Recall
Routine internal review of Device History Records identified six VITEK(R) 2 AST card lots were manufactured with less drug (gentamicin) than required for the Gentamicin High-Level Resistance (ghlr01n) screen test due to an error when making 1500mL and 2000mL media well preparations (solutions).
Firm Action
The firm initiated the recall by email on 08/13/2018. The firm requested the consignee discard the impacted product and contact the local bioMerieux representative for credit.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution US Nationwide and the countries of: Turkey, Austria, Switzerland, Czech Republic, Germany, Netherlands,
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.