Silver-Plated Copper Sims Uterine Sound, Part Number 64-600
FDA Recall Z-3188-2024 · Posted 2024-09-20
Advertisement
Recalling FirmCooperSurgical, Inc.
Recall StatusOpen, Classified
Product CodeHHM
Date Initiated2024-07-31
Product Quantity468 units
Reason for Recall
During the manufacturing process the devices were laser-marked with the incorrect part number.
Firm Action
An URGENT: MEDICAL DEVICE FIELD SAFETY NOTICE notification letter dated 7/31/24 was sent to customers.
Actions to be Taken:
For CUSTOMERS and DISTRIBUTORS:
" Inspect your inventory for any of this Product. Product Number and other information can be found on the box and package labeling.
" Please complete the included Customer Acknowledgment Form and send it via email to [email protected] to complete this action.
For CUSTOMERS:
" If your clinic or health system has multiple locations, and you wish to report for multiple physical locations, please indicate which physical locations in the response.
For DISTRIBUTORS:
" If the Product has been distributed to your customers, send a copy of this letter and the Customer Acknowledgement Form to each of these customers.
Please report any adverse events to your local Regulatory Authority as well as to CooperSurgical.
Our email address is [email protected]. You may also reach us at +1-203-601-5200 extension 3300 during normal operating hours of 09:00 17:00 M-F EDT.
Root Cause
Process control
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Armenia, Bahrain, Canada, Hong Kong, India, New Zealand, Qatar, Sweden.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.